Autors: Jekov I., Tsibulko V., Iliev, I. T. Title: ECG Database Applicable for Developmentand and Testing of Pace Detection Algorithms Keywords: Electrocardiogram, Pace pulses, ECG database, Pace detection Abstract: This paper presents an ECG database, named ‘PacedECGdb’ (available at http://biomed.bas.bg/bioautomation/2014/vol_18.4/files/PacedECGdb.zip), which contains different arrhythmias generated by HKP (Heidelberger Praxisklinik) simulator, combined with artificially superimposed pacing pulses that cover the wide ranges of rising edge (from <10 μs to 100 μs) and total pulse durations (from 100 μs to 2 ms) and correspond to various pacemaker modes. It involves a total number of 1404 recordings – 780 representing ‘pure’ ECG with pacing pulses and 624 that comprise paced ECGs corrupted by tremor. The signals are recorded with 9.81 μV/LSB amplitude resolution at 128 kHz sampling rate in order to preserve the steep raising and trailing edges of the pace pulses. References Issue
Copyright INT. J. BIOAUTOMATION Full text of the publication |
Цитирания (Citation/s):
1. Combined Filter for Rejection of 1st and 3rd Powerline Interference Harmonic from ECG signals - 2021 - в издания, индексирани в Scopus или Web of Science
2. Ivan Dotsinsky, An Approach to Successful Power-line Interference Suppression in ECG Signals, INT.J. BIOAUTOMATION, 2022, 26(1), 83-92 - 2022 - в издания, индексирани в Scopus или Web of Science
Вид: статия в списание, публикация в реферирано издание