Autors: Iliev, I. T., Tabakov, S. D., Tomchev, N. N. Title: An Analogue Front-end for Paced ECG Registration by Capacitive Electrodes Keywords: capacitive electrodes, pacemakers, remote monitoring Abstract: This paper presents the design of an analogue ECG front-end applicable for non-contact registration of cardiac activity using capacitive electrodes. While common instrumental amplifiers with such an application require extremely high input resistance, the proposed implementation is based on transimpedance amplifiers, allowing for a multifold reduction of the input resistance. The results of the LTSpice analysis show that the low cutoff frequency is 0.2 Hz (bellow than standard 0.5 Hz) as well as there is no distortion in the high frequency range, even when amplifying pacemaker pulses with approximately rectangular shape. References Issue
|
Цитирания (Citation/s):
1. Mihov G., Badarov D., Improved Adaptive Approach for Suppression of 1st and 3th Harmonic of Mains Interference in ECG Signals - 2022 - в издания, индексирани в Scopus или Web of Science
2. Mihov G., Badarov D., Improved Approach for Measuring Mains Interference - 2022 - в издания, индексирани в Scopus или Web of Science
3. Nikolova E., Dimitrov V., Hranov T, Gieva E., Remote Tracking System for ECG Signal with Textile Electrodes - 2022 - в издания, индексирани в Scopus или Web of Science
Вид: публикация в международен форум, публикация в издание с импакт фактор, публикация в реферирано издание, индексирана в Scopus